6.10 Alarm and Alert Audible Indicators
An audible indicator sounds when any of the following occurs:
•
A device inoperative condition occurs
•
The blower is on and a power failure occurs
•
An alarm condition occurs
•
An alert message appears on the screen
Alarm/Alert Type
Audible Indicator
Device Inoperative
When a device inoperative alarm occurs, a continuous audible indicator
sounds.
Power failure
When a power failure occurs, a series of beeps sounds in a 1 beep pat-
tern, repeating one second on, then one second off.
• •
High priority
When a high priority alarm is active, a series of beeps sounds in the fol-
lowing pattern, which is repeated twice: 3 beeps, a pause, and then 2
more beeps. This indicator continues until the cause of the alarm is cor-
rected or the audible alarm is silenced.
• • • • • • • • • •
Medium priority
When a medium priority alarm is active, a series of beeps sounds in a
3-beep pattern. This pattern repeats until the cause of the alarm is cor-
rected or the audible alarm is silenced. T
• • •
Alerts
When an alert appears on screen, a brief, 1-beep audible indicator
sounds.
•