To install the PCA:
1. Connect the DC and Blower Connectors to the PCA Connectors.
2. Seat the PCA onto the Blower Cap standoff and screw bosses, ensuring to position the Humidifier
Connector located on the PCA into the groove of the Rear Panel, and positioning the Flow and Pres-
sure Sensors into the Seals (refer to the illustrations below).
3. Secure the PCA with the two # 4 x 1/2” screws (torque to 2.5 IN-LB).
4. Assemble the remainder of the device as instructed in the previous sections.