P
AGE
137
1127179,
V
ER
.
02
D
REAM
S
TATION
B
I
PAP
S/T
AND
AVAPS
S
ERVICE
&
T
ECHNICAL
R
EFERENCE
M
ANUAL
Chapter 9.
Testing
Final testing of a device is a mandatory requirement after any repairs are made to the device, or if the device
enclosure has been opened for any reason.
NOTE
Testing must occur in a lab equipment environment only.
9.1 Required Equipment
Hardware:
• DreamStation Power Supply, 80W, (PN 1118499), Qty 1 ea.
• DreamStation Serial Accessory Module (PN 1123881 or PN 1126240 (includes serial and micro-USB ca-
bles listed below)), Qty 1 ea.
• Serial RS232 Cable DB9F-DB9M, 6 ft (PN 1037268), Qty 1 ea.
• USB Cable, Type A to Micro B (PN 1104765), Qty. 1 ea.
• Hose 22mm-18 in. (PN 1008198), Qty 1 ea.
• 15mm Heated Tube (PN HT15), Qty 1 ea.
• German Leak Device (ISO Port) (PN 1127131), Qty 1 ea.
• Pollen Filter (PN 1122446), Qty 1 ea.
• SD Card (PN 1063859 - 10 pk), Qty 1 ea.
• Printer (network or local connection)
• PC with the following minimum requirements:
• Windows 7 or Windows 10 32/64 Bit Operating System, Professional (Enterprise can be used for
Philips network users only)
• 3.20 GHz Processor
• 100 GB Free hard drive space
• 8 GB RAM
• 1 serial port
• Minimum 4 USB port
• Keyboard, mouse and 17 inch monitor
• Supports 220 VAC +/- 10% and 50 Hz +/- 1Hz, or 120 VAC +/- 10% and 60 Hz +/- 1Hz power