P
AGE
63
1127179,
V
ER
.
02
D
REAM
S
TATION
B
I
PAP
S/T
AND
AVAPS
S
ERVICE
&
T
ECHNICAL
R
EFERENCE
M
ANUAL
Code
Error Name
Error Level
Failed
Component
Actions
E-19
ERR_WDOG_TIMEOUT
ABORT
CPU
1. If error was NOT Last
Stop Error, clear error log
and test.
2. If error was Last Stop
Error, replace PCA and test.
E-20
ERR_MOTOR_SPINUP_F
LUX_LOW
REBOOT
1. Motor
connection
2. Blower Box
3. PCA
1. Reseat motor connector,
clear error log and test.
2. If error still occurs,
replace blower and retest.
3. If 2nd retest fails, replace
PCA.
E-21
ERR_MOTOR_VBUS_HI
GH
STOP
PCA
1. Replace PCA and retest
E-22
ERR_MOTOR_FLUX_MA
GNITUDE
REBOOT
1. Motor
connection
2. Blower Box
3. PCA
1. Reseat motor connector,
clear error log and test.
2. If error still occurs,
replace blower and retest.
3. If 2nd retest fails, replace
PCA.
E-23
ERR_MOTOR_OVERSPE
ED
REBOOT
PCA
1. If error was NOT Last
Stop Error, clear error log
and test.
2. If error was Last Stop
Error, replace PCA and test.
E-24
ERR_MOTOR_SPEED_R
EVERSE
REBOOT
PCA
1. If error was NOT Last
Stop Error, clear error log
and test.
2. If error was Last Stop
Error, replace PCA and test.