P
AGE
58
1127179,
V
ER
.
02
D
REAM
S
TATION
B
I
PAP
S/T
AND
AVAPS
S
ERVICE
&
T
ECHNICAL
R
EFERENCE
M
ANUAL
Available Functions for DreamStation Devices
Read SN & MN
This function allows you to read the serial and model
numbers of the device.
RTC Calibration
This function allows you to set and verify the real time clock
on the device.
Get Session ID
This function allows you to retrieve the session ID on the
device. The session ID is a unique number that interfaces
with Encore.
Set Session ID
This function allows you to set the session ID on the device.
This should only be executed when the Therapy PCA is
replaced on a device.
Read Time Meters
This function allows you to read the therapy and blower
hours on the device.
DS Logs (DreamStation
Logs)
This function allows you the retrieve certain data from the
device, including device error codes (if any are logged).
Within this function, you can also choose to clear the error
log, device log, therapy and/or blower hours, and first time
use. This function will also prompt you if the returned power
cord and power supply are functional, with the options of
“Yes”, “No”, or “Not Returned”. See Figure 5-4 below for an
example of data retrieved from the device.