Low Min Vent
Alarm
You can enable or disable this alarm by choosing 0 (off) to 99
lpm in 1.0 increments. The alarm will sound when the calcu-
lated minute ventilation is less than or equal to the specified
setting. The default is 0.
Low Tidal Vol-
ume
You can enable or disable this alarm by selecting On or Off.
This alarm will sound if IPAP max is reached and the target
tidal volume is not achieved. This alarm is only available when
AVAPS is enabled. The default is Off.