6.5 Device Error Codes
The following table lists the error level and descriptions for the DreamStation devices.
Error Level
Description
STOP
The error information is recorded in NVRAM and the unit is
placed into Safe State. The only functionality available to the
user is serial communication, turning off the audible alarm
via a key press and removing power.
REBOOT
The error information is recorded in NVRAM and the unit is
rebooted. The fifth occurrence of a REBOOT level error
within a 24 hour period (while power is maintained), will be
promoted by the system to a STOP level error.
ABORT
The error information is recorded in NVRAM. The fifth
occurrence of an ABORT level error within a 24 hour period
(while power is maintained), will be promoted by the system
to a STOP level error. This error is similar to a Reboot-Level
Error with the exception that the system is unable to handle
the error prior to the reset, e.g. watchdog timeout.
CONTINUE
The error information is recorded in NVRAM and the unit
continues to operate without noticeable alteration.
The following table should be used to aid in troubleshooting device error codes for the DreamStation devices.
Code
Error Name
Error Level
Failed
Component
Actions
E-0
ERR_NONE
STOP
N/A
N/A
E-3
ERR_INT_RAM
REBOOT
CPU
1. If error was NOT Last
Stop Error, clear error log
and test.
2. If error was Last Stop
Error, replace PCA and test.