Compressor Nebulizer 

 

Model UN-014 

Instruction Manual 

Original

 

EN

GL

ISH

 

Manuel d’instructions 

Traduction 

FRANÇAI

Manual de Instrucciones 

Traducción 

ESPA

ÑO

Manuale di Istruzioni 

Traduzione 

IT

AL

IANO

 

使用手冊

 

翻譯

 

中文

 

ﺕﺎﻣﻳﻠﻌﺗﻟﺍ ﻝﻳﻟﺩ

 

ﺔﻣﺟﺭﺗ

 

ﻲﺑ

ﺭﻋ

 

1WMPD4003398 

Summary of Contents for UN-014

Page 1: ...essor Nebulizer Model UN 014 Instruction Manual Original ENGLISH Manuel d instructions Traduction FRAN AIS Manual de Instrucciones Traducci n ESPA OL Manuale di Istruzioni Traduzione ITALIANO 1WMPD400...

Page 2: ...NTRODUCTION Page 5 6 SPECIFICATIONS Page 5 7 PARTS IDENTIFICATION Page 6 8 OPERATING INSTRUCTIONS Page 7 9 CLEANING Page 8 10 MAINTENANCE Page 8 11 TROUBLESHOOTING Page 9 12 IMPORTANT INFORMATION REGA...

Page 3: ...ry disorders Your Compressor Nebulizer should be used only as directed Do not use this device for anything other than as prescribed by your doctor Before use read this instruction manual thoroughly 2...

Page 4: ...scribed by the patient s doctor 4 This device is intended for aerosol therapy only Any other use is not recommended 5 Do not untwist the jet nebulizer during operation 6 Do not pour more than 6ml of s...

Page 5: ...ocked 9 Avoid operating in wet or damp locations 10 Unplug the unit before filling the nebulizer 11 When using this product near TVs microwave ovens pulse telephones X rays or other strong electric fi...

Page 6: ...o 50 psi 241 kpa to 345 kpa Neb Operating Pressure 9 16 psi 62 kpa to 110 kpa Liter Flow Range 5 8 L min Average Nebulization Rate Approx 0 25 mL min 6 ml medicine can be nebulized within 18 minutes M...

Page 7: ...lug UN 014A EC1B BF plug UN 014A EC2B A plug UN 014A EC3B O plug UN 014A EC4B Replacement parts list Appearance Part name Part code Qty Material Adult mask UNB 014A 01EC 1 PVC Child mask UNB 014A 02EC...

Page 8: ...lution through the opening on the cup using an eye dropper or pre measured dose container The 6cc marking on the nebulizer cup is the maximum level Medication cannot be filled beyond the maximum level...

Page 9: ...o need to clean the tube If necessary wipe the surface regularly Note Replace the nebulizer accessories including adult mask child mask mouthpiece nebulizer cup Air tube after using 6 months Replace t...

Page 10: ...of blockages The nebulizer kit is tilted at a sharp angle Make sure that the nebulizer kit is not tilted at an angle of more than 45 degrees The air tube is not connected correctly Make sure that the...

Page 11: ...cautions need to be observed Do not use mobile cellular telephones and medical devices which generate strong electrical or electromagnetic fields near the medical device This may result in incorrect o...

Page 12: ...Australasia Pty Ltd 32 Dew Street Thebarton South Australia 5031 AUSTRALIA Telephone 61 8 8301 8100 Fax 61 8 8352 7409 A D Instruments India Private Limited 509 Udyog Vihar Phase V Gurgaon 122 016 Har...