11. TROUBLESHOOTING 

Check the following if your unit should fail during operation. You can also refer 
to the pages of this manual for complete instructions. 

Problem 

Cause 

Remedy 

Nothing happens 
when the power 
switch is pressed. 

The AC adapter is not 
connected correctly to the 
electrical outlet or the 
compressor. 

Check that the plug is inserted in 
an electrical outlet. Unplug then 
reinsert the plug if necessary. 

No nebulization or 
low nebulization 
rate when the 
power is on. 

There is no medication in 
the medication tank. 

Add the correct amount of 
medication to the medication 
tank. 
 

There is too much/little 
medication in the 
medication tank. 
The vaporizer head is 
missing or not assembled 
correctly. 

Attach the vaporizer head 
correctly. 

The nebulizer kit is not 
assembled correctly. 

Assemble the nebulizer kit 
correctly. 

The nozzle is blocked. 

Make sure that the nozzle is free 
of blockages. 

The nebulizer kit is tilted at 
a sharp angle. 

Make sure that the nebulizer kit 
is not tilted at an angle of more 
than 45 degrees. 

The air tube is not 
connected correctly. 

Make sure that the air tube is 
correctly connected to the 
compressor and nebulizer kit. 

The air tube is folded or 
damaged. 

Make sure that the air tube does 
not contain kinks. 

The air tube is blocked. 

Make sure that the air tube is 
free of blockages. 

The air filter is dirty. 

Replace the air filter with a new 
one. 

The device is 
abnormally loud. 

The air filter cover is not 
attached correctly. 

Attach the air filter cover 
correctly. 

The device is very 
hot. 
 

The compressor is covered 
by something. 

Do not cover the compressor 
with any type of cover during 
use. 

The device was operated 
continuously for over 20 
minutes. 

Limit use to 20 minutes at a 
time, and allow a 40 minute rest 
interval before using the device 
again. 

 
NOTE: If the suggested remedy does not solve the problem, do not try to 

repair the device. No parts of the unit are user serviceable. 
Return the unit to an authorized retail outlet or distributor. 

Summary of Contents for UN-014

Page 1: ...essor Nebulizer Model UN 014 Instruction Manual Original ENGLISH Manuel d instructions Traduction FRAN AIS Manual de Instrucciones Traducci n ESPA OL Manuale di Istruzioni Traduzione ITALIANO 1WMPD400...

Page 2: ...NTRODUCTION Page 5 6 SPECIFICATIONS Page 5 7 PARTS IDENTIFICATION Page 6 8 OPERATING INSTRUCTIONS Page 7 9 CLEANING Page 8 10 MAINTENANCE Page 8 11 TROUBLESHOOTING Page 9 12 IMPORTANT INFORMATION REGA...

Page 3: ...ry disorders Your Compressor Nebulizer should be used only as directed Do not use this device for anything other than as prescribed by your doctor Before use read this instruction manual thoroughly 2...

Page 4: ...scribed by the patient s doctor 4 This device is intended for aerosol therapy only Any other use is not recommended 5 Do not untwist the jet nebulizer during operation 6 Do not pour more than 6ml of s...

Page 5: ...ocked 9 Avoid operating in wet or damp locations 10 Unplug the unit before filling the nebulizer 11 When using this product near TVs microwave ovens pulse telephones X rays or other strong electric fi...

Page 6: ...o 50 psi 241 kpa to 345 kpa Neb Operating Pressure 9 16 psi 62 kpa to 110 kpa Liter Flow Range 5 8 L min Average Nebulization Rate Approx 0 25 mL min 6 ml medicine can be nebulized within 18 minutes M...

Page 7: ...lug UN 014A EC1B BF plug UN 014A EC2B A plug UN 014A EC3B O plug UN 014A EC4B Replacement parts list Appearance Part name Part code Qty Material Adult mask UNB 014A 01EC 1 PVC Child mask UNB 014A 02EC...

Page 8: ...lution through the opening on the cup using an eye dropper or pre measured dose container The 6cc marking on the nebulizer cup is the maximum level Medication cannot be filled beyond the maximum level...

Page 9: ...o need to clean the tube If necessary wipe the surface regularly Note Replace the nebulizer accessories including adult mask child mask mouthpiece nebulizer cup Air tube after using 6 months Replace t...

Page 10: ...of blockages The nebulizer kit is tilted at a sharp angle Make sure that the nebulizer kit is not tilted at an angle of more than 45 degrees The air tube is not connected correctly Make sure that the...

Page 11: ...cautions need to be observed Do not use mobile cellular telephones and medical devices which generate strong electrical or electromagnetic fields near the medical device This may result in incorrect o...

Page 12: ...Australasia Pty Ltd 32 Dew Street Thebarton South Australia 5031 AUSTRALIA Telephone 61 8 8301 8100 Fax 61 8 8352 7409 A D Instruments India Private Limited 509 Udyog Vihar Phase V Gurgaon 122 016 Har...

Reviews: