English 5
5. INTRODUCTION
This compressor nebulizer is designed to deliver a prescribed medication
solution to treat patient respiratory disorders, such as asthma, allergies and
bronchitis. The nebulizer converts the medication solution into an aerosolized
mist which is inhaled by the patient through the mouthpiece or mask. Read this
manual thoroughly before using the nebulizer and store it for future reference.
6. SPECIFICATIONS
Electrical Requirements
230 V / 50 Hz
Power Consumption
Below 60 W
Rated Current
Below 0.70 A
Dimensions
Length 150 mm × Width 140 mm ×
Height 90 mm
Weight
1.2 kg (2.64 lbs)
Sound Pressure Level
55 dBA
Max. Compression Pressure
35 psi to 50 psi (241 kpa to 345 kpa)
Neb. Operating Pressure
9-16 psi (62 kpa to 110 kpa)
Liter Flow Range
5-8 L/min
Average Nebulization Rate:
Approx. 0.25 mL/min (6 ml medicine
can be nebulized within 18 minutes)
Maximum nebulizer solution capacity
6 ml
Maximum residual medicine volume
0.5 ml
Particle size (MMAD)
Approximately 3
μm
Operating Temperature/Humidity
+ 10 °C to + 40 °C,
30% RH to 85% RH
Storage/Transport Temperature
Range/ Humidity
- 20 °C to + 70 °C,
10% RH to 95% RH
Operating and Storage/Transport
Atmospheric Pressure
700 hpa to 1,060 hpa
Protection against electric shock
- Classified as Type
- Type BF applied parts: Mouthpiece, masks
Degree of safety in the presence of flammable anesthetics or oxygen —
No AP/APG (not suitable for use in the presence of flammable anesthetics or
oxygen)
Mode of operation — Continuous
IPX0 — No special protection against harmful ingress of water.