SERVICE MANUAL
Troubleshooting
I-Max Touch
8-51
8.5.
Detection and correction of possible defects in
dental radiographies
8.5.1.
Defects due to incorrect position of the patient
Problem
Description
Solution
Incisors too large and
blurred.
The patient is not positioned
correctly. He is too far back
from the optimal focal plane.
Position the patient correctly, and
verify that he is holding the bite with
his incisors on the fit notch and that
the bite holder rod is vertical.
Incisors too small and
blurred.
The patient is not positioned
correctly. He is too far
forward of the optimal focal
plane.
Position the patient correctly, and
verify that he is holding the bite with
his incisors on the fit notch and that
the bite holder rod is vertical.
Radiography with
blank central area.
The patient's spine inhibits
the passage of the X-rays
because it is too
compressed.
Check the alignment of the
Frankfurt plane. Try to stretch out
the cervical tract of the spine
correcting the height of the cursor
and moving the patient's feet
forward (see 6.4.3.3/4/6/7 of the
User Manual).
Dental arch not
symmetrical.
The median sagittal line
does not correspond to the
laser centering beam.
Realign the patient (see 6.4.3.6 of
the User Manual).
Upper apical area too
dark.
The patient does not keep
his lips closed and his
tongue is not against the
palate.
See paragraph 6.4.3.8 of the User
Manual.
Upper central apical
area out of focus.
The patient's head is turned
backwards (Frankfurt plane
not aligned).
Position the patient correcting the
Frankfurt plane.
The image is slanted
with respect to the
longitudinal axis of the
image and some of the
anatomical structures
are not symmetrical.
The patient's head is slanted
(not vertical).
Position the patient correcting the
sagittal plane.
The teeth are bigger on
one side than the
other.
The patient's head is turned
with respect to the axis of
the bite.
Position the patient correcting the
sagittal plane and checking that the
head is not turned.
(In CEPH examination)
white area on the
lower part of the film.
Panoramic chin-rest
mounted.
Repeat the examination removing
the PAN chin-rest.
Summary of Contents for i-max touch
Page 1: ...Version January 2014 Rev 4b I Max Touch 0051 Service Manual...
Page 2: ......
Page 27: ...SERVICE MANUAL Description I Max Touch 3 11 Figure 3 1 Block diagram...
Page 38: ...SERVICE MANUAL Description I Max Touch 3 22 THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 52: ...SERVICE MANUAL Technical data I Max Touch 4 14 THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 174: ...SERVICE MANUAL Schematics and drawings I Max Touch 9 2 THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 176: ......
Page 178: ......
Page 180: ......
Page 182: ......
Page 184: ......
Page 186: ......
Page 188: ......
Page 190: ......
Page 192: ......
Page 194: ......
Page 196: ......
Page 198: ......
Page 200: ......
Page 202: ......
Page 204: ......
Page 206: ......
Page 208: ......
Page 210: ......
Page 212: ......
Page 214: ......
Page 216: ......
Page 218: ......
Page 220: ......
Page 222: ......
Page 224: ......
Page 226: ......
Page 228: ......
Page 230: ......
Page 232: ......
Page 234: ......
Page 236: ......
Page 238: ......
Page 240: ......
Page 242: ......
Page 244: ......
Page 246: ......
Page 248: ......
Page 250: ...SERVICE MANUAL Spare Parts 10 2 THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 253: ...SERVICE MANUAL Spare Parts 10 5 2 4 8 3 6 5 7 12 13 14 16 1 9 10 11 15...
Page 257: ...SERVICE MANUAL Spare Parts 10 9 6 7 1 9 12 13 8 11 10 2 3 4 5 B8 B9 5 B6 B7...
Page 259: ...SERVICE MANUAL Spare Parts 10 11 8 6 7 1 B10 4 1 B11 2 5 3...
Page 261: ...SERVICE MANUAL Spare Parts 10 13 10 8 9 5 5 7 2 6 4 3 1 B11 6 4 1 B10...
Page 263: ...SERVICE MANUAL Spare Parts 4 6 5 7 8 9 1 2 3 10 11 12 6 14 13...
Page 265: ......