SERVICE MANUAL
Troubleshooting
I-Max Touch
8-11
8.2.3.
Errors with code from E300 to E399
8.2.3.1.
E300: Unexpected activation of zero position sensor on
secondary CEPH collimator /
E301: Timeout of zero position sensor on secondary CEPH
collimator /
E302: Zero position sensor for secondary CEPH collimator never
active /
E303: Zero position sensor for secondary CEPH collimator always
active
The X-ray beam coming out from the tube head assembly is collimated to
the area under exam by a secondary collimator that is moving
synchronously with the CEPH sensor holder; this collimator is moved by
a stepper motor M7.
All the above messages refer to this mechanism; all the considerations
above described for E280 to E283 for patient’s collision.
In case of absence of collision between the secondary collimator and the
patient, the cause can be a fault on motor M7 and/or optical sensor
barrier B11.
Check that during the power on phase there are movements on these
axes; the normal motion is going to engage the sensor and a reverse
motion to disengage it. If no motion is observed, the fault can be the
stepper motor or the cable connecting it to the main CPU (A5). Check
that the cables between the motor M7 and X4 of board A12 is well
inserted, as for the cable X1-X2 of A12 to X23, X24 and X34 of the
main CPU.
Check continuity of the above cables, from M7 to X4 A12 and from
X1-2 of A12 up to X23, X24 and X34 of the main CPU.
Repeat the power on sequence, verifying the motion.
If still there is no motion, replace the stepper motor and, if not fixed,
replace the CPU.
If a motion has been observed, but messages are displayed, this can be
the sensor or its wiring.
Power off the system and try to manually disengage the sensor,
taking care of a smooth motion.
Power on the system and check the functioning of the sensor,
measuring the DC voltage on the terminal of the sensor B11, with
and without an obstacle manually inserted. A variation of about 3.5
V minimum should be observed. If not, replace the sensor.
If the variation is observed and the message still continues to be
displayed, check the correct insertion of cable X3 on board A12 and
from X1-2 of the board up to X23, X24 and X34 of the main CPU;
check continuity of the interested wires and replace the faulty one (if
any).
If all cables are OK, the fault is on the main CPU, so replace it.
Summary of Contents for i-max touch
Page 1: ...Version January 2014 Rev 4b I Max Touch 0051 Service Manual...
Page 2: ......
Page 27: ...SERVICE MANUAL Description I Max Touch 3 11 Figure 3 1 Block diagram...
Page 38: ...SERVICE MANUAL Description I Max Touch 3 22 THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 52: ...SERVICE MANUAL Technical data I Max Touch 4 14 THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 174: ...SERVICE MANUAL Schematics and drawings I Max Touch 9 2 THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 176: ......
Page 178: ......
Page 180: ......
Page 182: ......
Page 184: ......
Page 186: ......
Page 188: ......
Page 190: ......
Page 192: ......
Page 194: ......
Page 196: ......
Page 198: ......
Page 200: ......
Page 202: ......
Page 204: ......
Page 206: ......
Page 208: ......
Page 210: ......
Page 212: ......
Page 214: ......
Page 216: ......
Page 218: ......
Page 220: ......
Page 222: ......
Page 224: ......
Page 226: ......
Page 228: ......
Page 230: ......
Page 232: ......
Page 234: ......
Page 236: ......
Page 238: ......
Page 240: ......
Page 242: ......
Page 244: ......
Page 246: ......
Page 248: ......
Page 250: ...SERVICE MANUAL Spare Parts 10 2 THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 253: ...SERVICE MANUAL Spare Parts 10 5 2 4 8 3 6 5 7 12 13 14 16 1 9 10 11 15...
Page 257: ...SERVICE MANUAL Spare Parts 10 9 6 7 1 9 12 13 8 11 10 2 3 4 5 B8 B9 5 B6 B7...
Page 259: ...SERVICE MANUAL Spare Parts 10 11 8 6 7 1 B10 4 1 B11 2 5 3...
Page 261: ...SERVICE MANUAL Spare Parts 10 13 10 8 9 5 5 7 2 6 4 3 1 B11 6 4 1 B10...
Page 263: ...SERVICE MANUAL Spare Parts 4 6 5 7 8 9 1 2 3 10 11 12 6 14 13...
Page 265: ......