SERVICE MANUAL
Technical data
I-Max Touch
4-9
4.3.
Measurement method of technical factors
WARNING:
These measurements require the removal of the HF group covers; this
means to gain access to internal parts where high voltage are normally
present.
For the measurement of the exposure parameters with the invasive
method, please follow the procedure described in paragraphs 7.2.1, 7.2.3
and 7.2.2 of this manual.
WARNING:
During the panoramic examination, the set value of kV and tube current
varies according to a pre-determined curve in order to compensate the
different absorption of X-ray beam due to different anatomical structures.
In this way, it is possible to obtain a good uniformity of the image
contrast. Particularly, the chosen value is lowered on the initial phase of
the panoramic and increased on the scissors/canine zone, in order to
compensate the effect of the cervical spine.
The value displayed during the panoramic examination corresponds to
the one chosen by the user, while the real value can be different; these
effects must be considered in case of measure of the exposure factors
using standard diagnostic mode. As an example, the variation follows the
curve hereafter:
Set value
Actual value kV
Actual value mA
Accuracy declared on the section "Technical data" is referred to the
actual value of kV and/or mA. In any case, the manufacturer guarantees
that the accuracy of the loading factors is always in compliance with the
international standard for safety of medical devices IEC 601-1.
Particularly, in accordance with IEC 601-2-7, the maximum deviation
(including the correction and instrument’s accuracy) is less than or equal
to
10 for kV, while for tube current is less than or equal to
15%.
Summary of Contents for i-max touch
Page 1: ...Version January 2014 Rev 4b I Max Touch 0051 Service Manual...
Page 2: ......
Page 27: ...SERVICE MANUAL Description I Max Touch 3 11 Figure 3 1 Block diagram...
Page 38: ...SERVICE MANUAL Description I Max Touch 3 22 THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 52: ...SERVICE MANUAL Technical data I Max Touch 4 14 THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 174: ...SERVICE MANUAL Schematics and drawings I Max Touch 9 2 THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 176: ......
Page 178: ......
Page 180: ......
Page 182: ......
Page 184: ......
Page 186: ......
Page 188: ......
Page 190: ......
Page 192: ......
Page 194: ......
Page 196: ......
Page 198: ......
Page 200: ......
Page 202: ......
Page 204: ......
Page 206: ......
Page 208: ......
Page 210: ......
Page 212: ......
Page 214: ......
Page 216: ......
Page 218: ......
Page 220: ......
Page 222: ......
Page 224: ......
Page 226: ......
Page 228: ......
Page 230: ......
Page 232: ......
Page 234: ......
Page 236: ......
Page 238: ......
Page 240: ......
Page 242: ......
Page 244: ......
Page 246: ......
Page 248: ......
Page 250: ...SERVICE MANUAL Spare Parts 10 2 THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 253: ...SERVICE MANUAL Spare Parts 10 5 2 4 8 3 6 5 7 12 13 14 16 1 9 10 11 15...
Page 257: ...SERVICE MANUAL Spare Parts 10 9 6 7 1 9 12 13 8 11 10 2 3 4 5 B8 B9 5 B6 B7...
Page 259: ...SERVICE MANUAL Spare Parts 10 11 8 6 7 1 B10 4 1 B11 2 5 3...
Page 261: ...SERVICE MANUAL Spare Parts 10 13 10 8 9 5 5 7 2 6 4 3 1 B11 6 4 1 B10...
Page 263: ...SERVICE MANUAL Spare Parts 4 6 5 7 8 9 1 2 3 10 11 12 6 14 13...
Page 265: ......