6 - 14
LLE11ARDA001A/E
6.13 Replacing the Ganged Water Supply Pipe AS
Replacement part: 40350-9300
1. Remove the SFB AS (40350-6000).
(See 6.12.)
2. Unscrew SB4×12 (n=2) and remove the cover
holder L (40350-M112). Previously
disconnect the cable.
3. Remove the hose connecting to the ganged
water supply pipe AS (40350-9300).
4. Unscrew H4×4 and remove the nozzle AS
(40350-9400).
5. Remove the pin (40350-M570) and ganged
water supply pipe AS (40350-9300).
6. Reassemble the parts in reverse order.