4 - 7
LLE11ARDA001A/E
4.3.1.1 Error 101-Roughing was not completed is displayed.
Is the voltage between 1-2 pins of the MD02 connector
of the CA44 cable (40350-CA44) DC+24V?
Does the CA17 cable (40350-CA17) break?
Replace the CA44 cable (40350-CA44) with the new
one.
Replace the CA17 cable (40350-CA17) with the new
one.
No
Yes
Yes
No
Is the symptom improved?
Does the Y axis move smoothly when checking its
operation by hand?
Replace the alfa step pulse motor for Y axis (40347-
E030) with the new one.
(See 6.23.2.)
Is the symptom improved?
Completion
Does the alfa step pulse motor (40347-E030) break?
Replace the Y axis AS (40350-5300) with the new one.
(See 6.17.)
Replace the Y axis AS (40350-5300) with the new one.
(See 6.17.)
Yes
No
Yes
No
No
Yes
Replace the BA02 board (40350-BA02) with the new
one.
(See 6.1.2.)
No
Is the voltage between 1-3(2-4) pins of the P902(J2)
c o n n e c t o r o n t h e B A 0 9 b o a r d ( 4 0 3 5 0 - B A 0 9 )
DC+24V?
Replace the CA09 cable (40350-CA09) with the new
one.
Yes
No
Is the voltage between 1-2, 3-4, 5-6, 7-8 pins of the
P909(J9) connector on the BA09 board (40350-BA09)
DC+24V?
Replace the BA09 board (40350-BA09) with the new
one.
(See 6.1.9.)
Yes
No
Completion
Yes
Dress the wheel.
(See 7.2.3.)
Is the parameter value of the maximum processing
rotation numbers 40?
Yes
No
Select 40 of the maximum processing rotation
numbers.
Summary of Contents for ME-1000
Page 8: ...1 2 LLE11ARDA001A E...
Page 12: ...3 2 LLE11ARDA001A E...
Page 120: ...7 34 LLE11ARDA001A E...
Page 121: ...8 1 LLE11ARDA001A E 8 SUPPLEMENT 8 1 Wiring Diagram...
Page 122: ...8 2 LLE11ARDA001A E 8 2 Connector Cable...
Page 123: ...8 3 LLE11ARDA001A E...
Page 125: ...8 5 LLE11ARDA001A E 8 4 Labels Top view Processing chamber...
Page 126: ...8 6 LLE11ARDA001A E Rear view or...