4 - 15
LLE11ARDA001A/E
Is the symptom improved?
Replace the BA02 board (40350-BA02) with the new
one.
(See 6.1.2.)
Copletion
Yes
No
Is the symptom improved?
Replace the BA01 board (40350-BA01) with the new
one.
(See 6.1.1.)
Completion
Yes
No
No
Perform the calibration of the lens measuring unit.
(See 7.3.)
Perform the adjustment of the feeler.
(See 7.4.)
Is the symptom improved?
Completion
Yes
No
4.3.2.5 205-Start & end measure data error is displayed.
Reassemble the feeler that rattles.
Does the left feeler (40350-M435) or the right feeler
(40350-M436) rattle?
4.3.2.4 The measurement data error is displayed.
Yes
No