4 - 12
LLE11ARDA001A/E
4.3.2.3 203-Feeler lift error is displayed.
Process the lens in EX mode.
Are the cataract lens, bifocal lens and such processed
in EX mode?
Is the voltage between 1-4 pins of the P204 (J4)
connector and 1-4 pins of the P205 (J5) connector on
the BA02 board (40350-BA02) DC+5V?
Does the CA13 cable (40350-CA13) break?
Is the voltage between 1-5 pins of the P705 (J5)
connector on the BA07 board (40350-BA07) of the left
lens measuring AS (40350-4100) and right lens
measuring AS (40350-4200) DC+5V?
Does the CA32 cable (40350-CA32) of the left lens
measuring AS (40350-4100) break?
Replace the BA02 board (40350-BA02) with the new
one.
(See 6.1.2.)
Replace the BA07 board (40350-BA07) of the left lens
measuring AS (40350-4100) and right lens measuring
AS (40350-4200) with the new one.
(See 6.1.7.)
Replace the CA32 cable (40350-CA32) of the left lens
measuring AS (40350-4100) with the new one.
No
Yes
No
Yes
No
Yes
No
Yes
No
Replace the CA13 cable (40350-CA13) with the new
one.
Yes
Does the left feeler (40350-M435) or the right feeler
(40350-M436) wear out remarkably?
Replace the left feeler (40350-M435) or the right feeler
(40350-M436) with the new one.
(See 6.18 and 6.19.)
No
Yes
Does the CA14 cable (40350-CA14) break?
Replace the CA14 cable (40350-CA14) with the new
one.
Yes
No
Does the CA32 cable (40350-CA32) of the right lens
measuring AS (40350-4200) break?
Replace the CA32 cable (40350-CA32) of the right
lens measuring AS (40350-4200) with the new one.
No
Yes
Establish the frame change mode.
Is the processing with frame change performed?
Yes
No
Summary of Contents for ME-1000
Page 8: ...1 2 LLE11ARDA001A E...
Page 12: ...3 2 LLE11ARDA001A E...
Page 120: ...7 34 LLE11ARDA001A E...
Page 121: ...8 1 LLE11ARDA001A E 8 SUPPLEMENT 8 1 Wiring Diagram...
Page 122: ...8 2 LLE11ARDA001A E 8 2 Connector Cable...
Page 123: ...8 3 LLE11ARDA001A E...
Page 125: ...8 5 LLE11ARDA001A E 8 4 Labels Top view Processing chamber...
Page 126: ...8 6 LLE11ARDA001A E Rear view or...