4 - 36
LLE11ARDA001A/E
Do the axes of the cup and lens shift in the processing?
Perform the axis adjustment.
(See 7.7.)
Use a protective tape in processing.
Yes
No
Is the symptom improved?
Use the soft processing mode.
Yes
No
Completion
4.8.3 Axis shift occurs.
4.8.4 Processing time is long.
Dress the wheel.
(See 7.2.3.)
Is the symptom improved?
Replace the wheel with the new one.
(See 6.2.)
Completion
No
Yes