3 - 1
LLE11ARDA001A/E
§3 TROUBLESHOOTING
Turn on the power switch.
Does the initialization start?
Does the starting screen appear?
Does any error message appear?
Can the lens measurement be performed properly?
Can the lens processing be performed properly?
Turn off the power switch.
Completion
4.2 The Initialization does not Start.
4.1 The Starting Screen does not Appear.
4.7 The Feeler does not Contact to the Lens.
4.8 The Lens Processing cannot be Performed
Properly.
Yes
Yes
Yes
No
Yes
No
No
No
No
4.3 The Error Message Appears.
Yes
Can the system be operated with the switches on the
control panel?
4.5 The System cannot be Operated with the Switches
on the Control Panel.
Yes
No
Does the maintenance message
“
Please Clean Tank &
Pump and Replace Filter
”
appear?
No
4.4 The Maintenance Message Appears.
Yes
Can the system be operated with the touch panel?
4.6 The System cannot be Operated with the Touch
Panel.
Yes
No
Summary of Contents for ME-1000