![Nidek Medical ME-1000 Service Manual Download Page 60](http://html1.mh-extra.com/html/nidek-medical/me-1000/me-1000_service-manual_1688027060.webp)
6 - 6
LLE11ARDA001A/E
6.2 Replacing the Grinding Wheel AS
Replacement part: 40350-1060
1. Open the top cover AS (40350-2000).
(See 5.1.)
2. Remove the hexagonal bolt (40340-M722) and
wheel stopper (40350-M174). Use the wheel
removing jig (40390-6000) to remove the
grinding wheel AS (40350-1060). Turn the
wheel inward to remove it since the screw
channel of the wheel is reverse.
3. Reassemble the parts in reverse order.
6.3 Replacing the Lens Clamp
Replacement part: 40370-3200
1. Open the top cover AS (40350-2000).
(See 5.1.)
2. Unscrew the flat point set screw (40370-M008)
and remove the lens clamp (40370-3200).
3. Reassemble the parts in reverse order.