4 - 33
LLE11ARDA001A/E
Replace the rotary encoder (40350-E035) of the left
lens measuring AS with the new one.
(See 6.21.1.)
Does the rotary encoder (40350-E035) of the left lens
measuring AS (40350-4100) break?
4.3.3.8 Error code 408-Front feeler initialize error is
displayed.
Yes
No
Does the feeler slide smoothly in the X axis direction?
Replace the left and right lens measuring units (40350-
4100, 4200) with the new one.
(See 6.9.)
No
Yes
4.7 The Feeler does not Touch the Lens.
4.7.1 The front feeler does not touch the lens.
4.7.2 The rear feeler does not touch the lens.
Replace the rotary encoder (40350-E035) of the right
lens measuring AS (40350-4200) with the new one.
(See 6.21.2.)
Does the rotary encoder (40350-E035) of the right lens
measuring AS (40350-4200) break?
4.3.3.9 Error code 409-Rear feeler Initialize error is
displayed.
Yes
No
Does the feeler slide smoothly in the X axis direction?
Replace the left and right lens measuring units (40350-
4100, 4200) with the new one.
(See 6.9.)
No
Yes