5.9 Error correction
5
-
48
FRA 51615
12) Set the load standard value format. Use
【
Calibration
】
-
[LOAD CORRECTION]
-
[LOAD
STANDARD FORMAT]
to select the standard value format from among
[Cs-D]
,
[Cp-D]
,
[Rp-Cp]
,
[Rs-Ls]
,
[Rs-X]
, and
[Z-
θ]
.
13) Enter the load standard value. Select
【
Calibration
】
-
[Load Standard Value]
and press the
ENTER
key. In the table that is displayed, enter the load standard values. The entered load standard
values must be within the measurement frequency range and be entered for at least three frequencies.
Values can be entered for up to 10 frequencies.
After entering the load standard values, press the
CANCEL
key to close the value input table.
14) Set
【
Calibration
】
-
[LOAD CORRECTION]
to
[ON]
. Load correction can be done normally with
-
[OPEN CORRECTION]
and
[SHORT CORRECTION]
set to
[OFF]
.
The preparation for load correction is completed by performing the steps described above. If you
disconnect the standard impedance device and perform the measurement with the DUT connected to the
measurement ports, load-corrected measurement data can be obtained.
The correction data that has been copied to the open, short, and load correction memory is retained even if
the power to the instrument is turned off, so it is available for use the next time the power is turned on.
However, these functions are also executed a part of the error correction process of the instrument itself, so
we recommend that the open, short, and load correction data be measured again after the internal
temperature has stabilized (about 30 minutes after power up) and after any change in the ambient
temperature.
Summary of Contents for FRA51615
Page 2: ......
Page 3: ...FREQUENCY RESPONCE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS DA00061633 001 ...
Page 24: ...1 3 Operating principles 1 10 FRA51615 Blank ...
Page 31: ...2 2 Installation 2 7 FRA51615 Figure 2 1 Rack mounting kit assembly diagram JIS ...
Page 32: ...2 2 Installation 2 8 FRA51615 Figure 2 2 Rack mounting kit assembly diagram EIA ...
Page 33: ...2 2 Installation 2 9 FRA51615 Figure 2 3 Rack mounting kit dimensions JIS ...
Page 34: ...2 2 Installation 2 10 FRA51615 Figure 2 4 Rack mounting kit dimensions EIA ...
Page 134: ...3 6 Initial settings 3 96 FRA51615 Blank ...
Page 284: ...9 9 Miscellaneous specifications 9 20 FRA51615 Figure 9 3 External dimensions ...
Page 300: ...XVI FRA51615 Blank ...
Page 302: ......
Page 304: ......
Page 305: ......