1.2 Applications
1
-
5
FRA 51615
■ Example of impedance characteristic measurement
The impedance of the DUT can be measured by using this instrument to measure the voltage (v) with CH1
and the current (i) with CH2, and calculating the impedance by taking the vector ratio of CH1 and CH2.
The current of the DUT (i) is converted to a voltage with a shunt resistance, current probe, or CT, etc., and
the voltage is input to CH2.
Figure 1-3
Example of impedance characteristic measurement
It is possible to measure the impedance of the DUT at high voltage or large current by amplification of the
output of the sweep oscillator with the power amplifier as needed. This instrument differs from LCR meters,
impedance analyzers, and other such devices that limit the signal level to less than 1 V in that measurement
is possible under the signal amplitude conditions actually used by the EUT, thus enabling evaluation of
materials that handle high energy levels, such as magnets and dielectrics.
v
POWER AMPLIFIER
(OPTION)
i
CT
CH2
OSC
CH1
Z=v/i=V1/V2 x K
CT
( K
CT
=V2/i )
V1
V2
DUT
(Z)
Summary of Contents for FRA51615
Page 2: ......
Page 3: ...FREQUENCY RESPONCE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS DA00061633 001 ...
Page 24: ...1 3 Operating principles 1 10 FRA51615 Blank ...
Page 31: ...2 2 Installation 2 7 FRA51615 Figure 2 1 Rack mounting kit assembly diagram JIS ...
Page 32: ...2 2 Installation 2 8 FRA51615 Figure 2 2 Rack mounting kit assembly diagram EIA ...
Page 33: ...2 2 Installation 2 9 FRA51615 Figure 2 3 Rack mounting kit dimensions JIS ...
Page 34: ...2 2 Installation 2 10 FRA51615 Figure 2 4 Rack mounting kit dimensions EIA ...
Page 134: ...3 6 Initial settings 3 96 FRA51615 Blank ...
Page 284: ...9 9 Miscellaneous specifications 9 20 FRA51615 Figure 9 3 External dimensions ...
Page 300: ...XVI FRA51615 Blank ...
Page 302: ......
Page 304: ......
Page 305: ......