5.9 Error correction
5
-
44
FRA 51615
9) Enable the correction function. Use
【
Measure
】
-
[OPEN CORRECTION]
or
-
[SHORT
CORRECTION]
to set the correction function to ON.
The preparation for open correction or short correction is completed by performing the steps described
above. Measurement data that has been corrected for stray admittance or residual impedance in the
measurement system can be obtained by connecting the DUT to the measurement ports and performing the
measurements.
The correction data that has been copied to the open correction memory or the short correction memory is
retained, even if the power to the instrument is turned off, so it is available for use the next time the power
is turned on. However, these functions are also executed as a part of the error correction process of the
instrument itself, so we recommend that the open correction data or short correction data be measured again
after the internal temperature has stabilized (about 30 minutes after power up) and after any change in the
ambient temperature.
Summary of Contents for FRA51615
Page 2: ......
Page 3: ...FREQUENCY RESPONCE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS DA00061633 001 ...
Page 24: ...1 3 Operating principles 1 10 FRA51615 Blank ...
Page 31: ...2 2 Installation 2 7 FRA51615 Figure 2 1 Rack mounting kit assembly diagram JIS ...
Page 32: ...2 2 Installation 2 8 FRA51615 Figure 2 2 Rack mounting kit assembly diagram EIA ...
Page 33: ...2 2 Installation 2 9 FRA51615 Figure 2 3 Rack mounting kit dimensions JIS ...
Page 34: ...2 2 Installation 2 10 FRA51615 Figure 2 4 Rack mounting kit dimensions EIA ...
Page 134: ...3 6 Initial settings 3 96 FRA51615 Blank ...
Page 284: ...9 9 Miscellaneous specifications 9 20 FRA51615 Figure 9 3 External dimensions ...
Page 300: ...XVI FRA51615 Blank ...
Page 302: ......
Page 304: ......
Page 305: ......