![NF FRA51615 Instruction Manual Download Page 275](http://html1.mh-extra.com/html/nf/fra51615/fra51615_instruction-manual_1684898275.webp)
9.4 Analysis processing section
9
-
11
FRA 51615
Inter-range accuracy
Upper: gain (ratio); Middle: impedanse Z; Lower: phase
Measurement range
(rms)
Frequency
≦
100 kHz
≦
200 kHz
≦
2 MHz
≦
15 MHz
600 V
-
±0.1 dB
±1.2 %
±0.6°
-
-
-
300 V
±0.05 dB
±0.58 %
±0.3°
±0.1 dB
±1.2 %
±0.6°
-
-
100 V
±0.05 dB
±0.58 %
±0.3°
-
-
30 V
±0.05 dB
±0.58 %
±0.3°
-
10 V
±0.05 dB
±0.58 %
±0.3°
3 V
±0.05 dB
±0.58 %
±0.3°
1 V
±0.05 dB
±0.58 %
±0.3°
300 mV
±0.05 dB
±0.58 %
±0.3°
100 mV
-
Conditions: At least 30 cycles of integration and fixed measurement range for both channels.
The gain and phase error for when the measurement range difference between channels
is 1, the input signal levels of both channels are equal, and equal to the range full scale
level of the smaller range.
*For the cells that contain only " – ", the measurement range exceeds the maximum input voltage (see
"
9.2 Analysis input section
, ● Maximum input voltage"), so either measurement is not possible or
there is no accuracy specification for it.
Summary of Contents for FRA51615
Page 2: ......
Page 3: ...FREQUENCY RESPONCE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS DA00061633 001 ...
Page 24: ...1 3 Operating principles 1 10 FRA51615 Blank ...
Page 31: ...2 2 Installation 2 7 FRA51615 Figure 2 1 Rack mounting kit assembly diagram JIS ...
Page 32: ...2 2 Installation 2 8 FRA51615 Figure 2 2 Rack mounting kit assembly diagram EIA ...
Page 33: ...2 2 Installation 2 9 FRA51615 Figure 2 3 Rack mounting kit dimensions JIS ...
Page 34: ...2 2 Installation 2 10 FRA51615 Figure 2 4 Rack mounting kit dimensions EIA ...
Page 134: ...3 6 Initial settings 3 96 FRA51615 Blank ...
Page 284: ...9 9 Miscellaneous specifications 9 20 FRA51615 Figure 9 3 External dimensions ...
Page 300: ...XVI FRA51615 Blank ...
Page 302: ......
Page 304: ......
Page 305: ......