5.5 Graph display
5
-
17
FRA 51615
■ The unwrapped phase display function
For the
[UNWRAP]
display setting, unwrapped phase display is performed, with the phase measurement
value at the sweep starting frequency (the lower-limit frequency for an upsweep or the upper-limit
frequency for a downsweep) as the reference value. Thus, when DUTs that have entirely the same phase
characteristics are measured by upsweep or downsweep measurement, the results are displayed in graphs
that differ in phase by an integer multiple of 360°.
■ The 360° phase shift function
To display the currently-displayed phase data as a graph with 360° added or subtracted, select
【
Graph
】
–
[PHASE RANGE]
–
[SHIFT]
and tap on either of the function keys labeled
[+360°]
or
[
−360°]
. This is
used when a phase value in a phase continuous display is too large. When the phase display range setting
is anything other than
[UNWRAP]
, the display is not changed.
2) Tap
1) Select SHIFT
Summary of Contents for FRA51615
Page 2: ......
Page 3: ...FREQUENCY RESPONCE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS DA00061633 001 ...
Page 24: ...1 3 Operating principles 1 10 FRA51615 Blank ...
Page 31: ...2 2 Installation 2 7 FRA51615 Figure 2 1 Rack mounting kit assembly diagram JIS ...
Page 32: ...2 2 Installation 2 8 FRA51615 Figure 2 2 Rack mounting kit assembly diagram EIA ...
Page 33: ...2 2 Installation 2 9 FRA51615 Figure 2 3 Rack mounting kit dimensions JIS ...
Page 34: ...2 2 Installation 2 10 FRA51615 Figure 2 4 Rack mounting kit dimensions EIA ...
Page 134: ...3 6 Initial settings 3 96 FRA51615 Blank ...
Page 284: ...9 9 Miscellaneous specifications 9 20 FRA51615 Figure 9 3 External dimensions ...
Page 300: ...XVI FRA51615 Blank ...
Page 302: ......
Page 304: ......
Page 305: ......