5.9 Error correction
5
-
50
FRA 51615
under those conditions is NaN (Not a Number).
•
The load standard values and the load correction data are stored in different memories within the
instrument. The load standard values are stored in the measurement conditions memory and the load
correction data are stored in a dedicated memory. For that reason, if the measurement conditions are
restored, the load correction data remains as it is, but the load standard values are overridden by the
restored measurement conditions. If the load standard values that are in the restored measurement
conditions do not correspond to the load correction data, the load correction results will be incorrect.
Therefore, take appropriate care when restoring the measurement conditions.