3.5 Operation screen reference
3
-
49
FRA 51615
■ [START DELAY]
: Set the delay for starting measurement.
This sets the delay time (or number of cycles) for the beginning of a sweep measurement or a spot
measurement. This is set for cases in which a large change in frequency generates a transient response in
the circuit under test. For information on delays, refer to "
5.2 Delay
".
○ Setting method
< Numerical value settings (immediate setting) > , < Numerical value settings (new input) > ,
< Numerical value settings (edit) >
For time, the setting range is 0 to 9,990 seconds and the resolution is 3 digits or 0.1 ms, whichever is
larger. For cycles, the range is 0 to 9,999 cycles and the resolution is one cycle.
○ Function key functions
• [DIRECT NEW]
: Set the start delay with the numerical input dialog
【
NUM
】
.
• [DIRECT EDIT]
: Edit the start delay with the numerical input dialog
【
NUM
】
.
○ Functions of the function keys set with the numerical input dialog
【
NUM
】
• [CYCLE]
: Set the entered value in units of cycle.
• [msec]
: Set the entered value in units of milliseconds.
• [sec]
: Set the entered value in units of seconds.
■ [MEAS DELAY]
: Set the delay for measurement.
This sets the delay time (or cycles) for the beginning of a sweep measurement after a frequency change
occurs during a sweep. This is set for measurement of circuits that have sudden changes in frequency. For
information on delays, refer to "
5.2 Delay
".
○ Setting method
< Numerical value settings (immediate setting) > , < Numerical value settings (new input) > ,
< Numerical value settings (edit) >
For time, the setting range is 0 to 9,990 seconds and the resolution is 3 digits or 0.1 ms, whichever is
larger. For cycles, the range is 0 to 9,999 cycles and the resolution is one cycle.
○ Function key functions
• [DIRECT NEW]
: Set the measurement delay with the numerical input dialog
【
NUM
】
.
• [DIRECT EDIT]
: Edit the measurement delay with the numerical input dialog
【
NUM
】
.
○ Functions of the function keys set with the numerical input dialog
【
NUM
】
• [CYCLE]
: Set the entered value in units of cycles.
• [msec]
: Set the entered value in units of milliseconds.
• [sec]
: Set the entered value in units of seconds.
Summary of Contents for FRA51615
Page 2: ......
Page 3: ...FREQUENCY RESPONCE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS DA00061633 001 ...
Page 24: ...1 3 Operating principles 1 10 FRA51615 Blank ...
Page 31: ...2 2 Installation 2 7 FRA51615 Figure 2 1 Rack mounting kit assembly diagram JIS ...
Page 32: ...2 2 Installation 2 8 FRA51615 Figure 2 2 Rack mounting kit assembly diagram EIA ...
Page 33: ...2 2 Installation 2 9 FRA51615 Figure 2 3 Rack mounting kit dimensions JIS ...
Page 34: ...2 2 Installation 2 10 FRA51615 Figure 2 4 Rack mounting kit dimensions EIA ...
Page 134: ...3 6 Initial settings 3 96 FRA51615 Blank ...
Page 284: ...9 9 Miscellaneous specifications 9 20 FRA51615 Figure 9 3 External dimensions ...
Page 300: ...XVI FRA51615 Blank ...
Page 302: ......
Page 304: ......
Page 305: ......