5.9 Error correction
5
-
39
FRA 51615
5) When the sweep measurement is completed, select
【
Trace
】
–
[MEAS COPY]
and tap the
[EQUALIZE CORR MEM]
function key to copy the measurement data to the internal equalizing
memory of the instrument.
6) Set
【
Calibration
】
–
[EQUALIZING]
to
[ON]
.
7) Restore the original connections as they were prior to removing the EUT in step 3 ("
Figure 5-20
Example of connections for gain-phase measurement
").
The procedure described above completes the equalizing preparation. Performing a sweep measurement
produces measurement data that is corrected for error in probes, cables and other such parts of the
measurement system.
The correction data that has been copied to the equalizing memory is retained, even if the instrument power
is turned off. It is therefore available and ready for use the next time the power is turned on. However, this
function is executed as part of the error correction process for the instrument itself, so we recommend that
the equalizing correction data be measured again when the internal temperature of the instrument has
stabilized after the power is turned on (about 30 minutes) or when there has been a sudden change in the
ambient temperature, etc.
Summary of Contents for FRA51615
Page 2: ......
Page 3: ...FREQUENCY RESPONCE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS DA00061633 001 ...
Page 24: ...1 3 Operating principles 1 10 FRA51615 Blank ...
Page 31: ...2 2 Installation 2 7 FRA51615 Figure 2 1 Rack mounting kit assembly diagram JIS ...
Page 32: ...2 2 Installation 2 8 FRA51615 Figure 2 2 Rack mounting kit assembly diagram EIA ...
Page 33: ...2 2 Installation 2 9 FRA51615 Figure 2 3 Rack mounting kit dimensions JIS ...
Page 34: ...2 2 Installation 2 10 FRA51615 Figure 2 4 Rack mounting kit dimensions EIA ...
Page 134: ...3 6 Initial settings 3 96 FRA51615 Blank ...
Page 284: ...9 9 Miscellaneous specifications 9 20 FRA51615 Figure 9 3 External dimensions ...
Page 300: ...XVI FRA51615 Blank ...
Page 302: ......
Page 304: ......
Page 305: ......