3.5 Operation screen reference
3
-
47
FRA 51615
■ [ Amplitude Compression ]
–
[CORRECTION FACTOR]
: Set the correction factor.
This sets the correction factor (control sensitivity) that is used in amplitude adjustment. Usually, 100%
works well, but the stability of control may depend on the system being measured. If instability is
observed, reduce the correction factor.
○ Setting method
< Numerical value settings (immediate setting) > , < Numerical value settings (new input) > ,
< Numerical value settings (edit) >
The setting range is 1 to 100% and the resolution is 1%.
○ Function key functions
• [DIRECT NEW]
: Set the correction factor with the numerical input dialog
【
NUM
】
.
• [DIRECT EDIT]
: Edit the correction factor with the numerical input dialog
【
NUM
】
.