3.2 Input and output connectors
3
-
9
FRA 51615
3.2 Input and output connectors
■ Analyzer input connectors (CH1 and CH2)
The input and output channels of the instrument analysis module (
CH1
and
CH2
) are each electrically
isolated from the instrument enclosure. The oscillator output (
OSC
) is also isolated.
Maximum input voltage : 600 V CAT II or 300 V CAT III (BNC signal to BNC ground)
Input impedance
: 1 M Ω ±2%, 20 pF ±5 pF in parallel
Isolation voltage
: 600 V CAT II or 300 V CAT III (BNC ground to enclosure)
Connector
: Front panel, insulated BNC connector
The isolation of the input connectors and the enclosure is maintained even when the power is turned off.
Figure 3-3
Isolation rating of the analyzer input connectors
!
WARNING
Do not connect this instrument to circuits that exceed MEASUREMENT CATEGORY
II/600 V or MEASUREMENT CATEGORY III/300 V. Doing so may result in insulation
breakdown and risk of electrical shock. This instrument may also be damaged by
connection to circuits that exceed the specifications.
1
MΩ
CH1
1
ATT
1
MΩ
CH2
2
ATT
Enclosure
CH1
600 V CAT
Ⅱ
300 V CAT
Ⅲ
600 V CAT
Ⅱ
300 V CAT
Ⅲ
600 V CAT
Ⅱ
300 V CAT
Ⅲ
CH2
600 V CAT
Ⅱ
300 V CAT
Ⅲ
600 V CAT
Ⅱ
300 V CAT
Ⅲ
600 V CAT
Ⅱ
300 V CAT
Ⅲ
Summary of Contents for FRA51615
Page 2: ......
Page 3: ...FREQUENCY RESPONCE ANALYZER FRA51615 INSTRUCTION MANUAL OPERATIONS DA00061633 001 ...
Page 24: ...1 3 Operating principles 1 10 FRA51615 Blank ...
Page 31: ...2 2 Installation 2 7 FRA51615 Figure 2 1 Rack mounting kit assembly diagram JIS ...
Page 32: ...2 2 Installation 2 8 FRA51615 Figure 2 2 Rack mounting kit assembly diagram EIA ...
Page 33: ...2 2 Installation 2 9 FRA51615 Figure 2 3 Rack mounting kit dimensions JIS ...
Page 34: ...2 2 Installation 2 10 FRA51615 Figure 2 4 Rack mounting kit dimensions EIA ...
Page 134: ...3 6 Initial settings 3 96 FRA51615 Blank ...
Page 284: ...9 9 Miscellaneous specifications 9 20 FRA51615 Figure 9 3 External dimensions ...
Page 300: ...XVI FRA51615 Blank ...
Page 302: ......
Page 304: ......
Page 305: ......