8.5 Performance test
8-6
FRA 51615
8.5.2 Oscillator frequency accuracy
Before performing each of the following measurements, initialize this instrument.
This instrument
: Set the frequency to 1.0 MHz.
Measuring instrument
: Use a frequency counter and set the gate time to 10 seconds.
Connections
: Connect the
OUTPUT OSC
connector of this instrument to the
frequency counter with a coaxial cable.
Measurement method
: Read the measured value from the frequency counter.
Judgment
Passing range
: 0.999990 kHz to 1.000010 MHz
Measured value
: _ ._ _ _ _ _ _ MHz
FRA 51615
OSC
CH1
CH2
BNC-BNC
coaxial cable
Frequency counter