![NF EC1000SA Instruction Manual Download Page 276](http://html1.mh-extra.com/html/nf/ec1000sa/ec1000sa_instruction-manual_1684880276.webp)
9.14
Withstand Voltage and Insulation Resistance
EC1000SA
9-16
9.14 Withstand Voltage and Insulation Resistance
Between input power line vs. output/chassis or input power line/ chassis vs. output
a) Withstand voltage
AC 1500 V, 1 minute
b) Insulation resistance
30 MΩ or higher, 500 V DC
9.15 Safety, EMC and RoHS
Only Models with a CE Marking on the Rear Panel.
a) Safety
Compliant with the following standard requirement
EN61010-1
Pollution Degree 2
b) EMC
Compliant with the following standard requirement
EN61326-1 (Group 1, Class A)
EN61000-3-2
EN61000-3-3
Note:
This product may cause interference if used in residential
areas. Such use must be avoided unless the user takes
special measures to reduce electromagnetic emissions to
prevent interference to the reception of radio and
television broadcasts.
c) RoHS
Compliant with the following
Directive 2011/65/EU