![NF EC1000SA Instruction Manual Download Page 201](http://html1.mh-extra.com/html/nf/ec1000sa/ec1000sa_instruction-manual_1684880201.webp)
6.3
Command Descriptions
EC1000SA
6-23
[SOURce:]SEQuence:CSTep?
This returns the step number being executed. The response ranges from
−
1 and 1 to 255.
−
1 is
returned in the idle mode.
Syntax:
[SOURce:]SEQuence:CSTep?
Parameters:
None
Query example:
SEQ:CST?
Response format:
<NR1>
Response example:
2
[SOURce:]SEQuence:LEN?
This returns the maximum number of sequence steps. 255 is always returned.
Syntax:
[SOURce:]SEQuence:LEN?
Parameters:
None
Query example:
SEQ:LEN?
Response format:
<NR1>
Response example:
255
[SOURce:]SEQuence:COUNt?
This returns the jump times of the step specified by SEQ:STEP. Response range of the number of
the jump times is -1 to 999. In the idle mode,
−
1 is returned.
Syntax:
[SOURce:]SEQuence:COUNt?
Parameters:
None
Query example:
SEQ:COUN?
Response format:
<NR1>
Response example:
2