6.3
Command Descriptions
EC1000SA
6-35
MEASure[:SCALar]:VOLTage:HIGH?
This returns the maximum peak value (AC+DC peak value) of the output voltage in Vpk units. For
the 100 V range, the measurement full scale of the maximum peak value of output voltage is ±250
Vpk, and for the 200 V range, it is ±500 Vpk. If the measured value exceeds the measurement
range, 999 Vpk is returned as a measurement error.
Syntax:
MEASure[:SCALar]:VOLTage:HIGH?
Parameters:
None
Query example:
MEAS:VOLT:HIGH?
Response format:
<NR1>
Response example:
100
MEASure[:SCALar]:VOLTage:LOW?
This returns the minimum peak value (AC+DC peak value) of the output voltage in Vpk units. For
the 100 V range, the measurement full scale of the minimum peak value of output voltage is ±250
Vpk, and for the 200 V range, it is ±500 Vpk. If the measured value exceeds the measurement
range,
−
999 Vpk is returned as a measurement error.
Syntax:
MEASure[:SCALar]:VOLTage:LOW?
Parameters:
None
Query example:
MEAS:VOLT:LOW?
Response format:
<NR1>
Response example:
−
99
MEASure[:SCALar]:VOLTage:AVErage?
This returns the output voltage average value (AC+DC average value) in V units. For the 100 V
range, the measurement full scale of the average value of output voltage is ±250.0 V, and for the
200 V range, it is ±500.0 V. If the measured value exceeds the measurement range, 999.9 V is
returned as a measurement error.
Syntax:
MEASure[:SCALar]:VOLTage:AVErage?
Parameters:
None
Query example:
MEAS:VOLT:AVE?
Response format:
<NR2>
Response example:
150.2
Summary of Contents for EC1000SA
Page 1: ...NF Corporation PROGRAMMABLE AC DC POWER SOURCE EC1000SA Instruction Manual nbn Austria GmbH...
Page 2: ...DA00036667 003 PROGRAMMABLE AC DC POWER SOURCE EC1000SA Instruction Manual...
Page 236: ...6 9 Notes on Remote Control EC1000SA 6 58...
Page 260: ...8 5 Backup Battery EC1000SA 8 8...
Page 280: ...9 17 External Dimensions and Weight EC1000SA 9 20...