![NF EC1000SA Instruction Manual Download Page 216](http://html1.mh-extra.com/html/nf/ec1000sa/ec1000sa_instruction-manual_1684880216.webp)
6.3
Command Descriptions
EC1000SA
6-38
STATus:OPERation:NTRansition
STATus:OPERation:PTRansition
This sets the operation transition filter (OPTF). The setting ranges from 0 to 20482.
•
When the NTR filter is set to 1, OPER is set to 1 when OPCR changes from 1 to 0.
•
When the PTR filter is set to 1, OPER is set to 1 when OPCR changes from 0 to 1.
•
When the NTR and PTR filters are both set to 1, OPER is set to 1 when OPCR changes.
•
When the NTR and PTR filters are both set to 0, changes in OPCR are not propagated to OPER.
Syntax:
STATus:OPERation:NTRansition
<NR1>
STATus:OPERation:PTRansition
<NR1>
Parameters:
0 to 20482
Command example:
STAT:OPER:NTR 16384
STAT:OPER:PTR 0
Query example:
STAT:OPER:NTR?
STAT:OPER:PTR?
Response format:
<NR1>
Response example:
16384 (when NTR is queried)
0 (when PTR is queried)
STATus:WARNing:CONDition?
The warning condition register (WRCR) value is returned. The response ranges from 0 to 29947.
Syntax:
STATus:WARNing:CONDition?
Parameters:
None
Query example:
STAT:WARN:COND?
Response format:
<NR1>
Response example:
16384
Summary of Contents for EC1000SA
Page 1: ...NF Corporation PROGRAMMABLE AC DC POWER SOURCE EC1000SA Instruction Manual nbn Austria GmbH...
Page 2: ...DA00036667 003 PROGRAMMABLE AC DC POWER SOURCE EC1000SA Instruction Manual...
Page 236: ...6 9 Notes on Remote Control EC1000SA 6 58...
Page 260: ...8 5 Backup Battery EC1000SA 8 8...
Page 280: ...9 17 External Dimensions and Weight EC1000SA 9 20...