![NF EC1000SA Instruction Manual Download Page 225](http://html1.mh-extra.com/html/nf/ec1000sa/ec1000sa_instruction-manual_1684880225.webp)
6.3
Command Descriptions
EC1000SA
6-47
∗∗∗∗
RST
This resets the device to its factory settings. However, reset cannot be executed when output is on.
([1, “Invalid with output on”] error occurs.)
Syntax:
∗∗∗∗
RST
Parameters:
None
Command example:
∗
RST
Response format:
None
Response example:
None
∗∗∗∗
SAV
This stores the current status information in memory. The memory storage ranges from 1 to 30.
Syntax:
∗∗∗∗
SAV
<NR1>
Parameters:
1 to 30
Command example:
∗
SAV 1
Response format:
None
Response example:
None
∗∗∗∗
SRE
This sets the service request enable register. The setting ranges from 0 to 178.
Syntax:
∗∗∗∗
SRE
<NR1>
Parameters:
0 to 178
Command example:
∗
SRE 178
Query example:
∗
SRE?
Response format:
<NR1>
Response example:
178
∗∗∗∗
STB?
This returns the status byte register value. The response ranges from 0 to 178.
Syntax:
∗∗∗∗
STB?
Parameters:
None
Query example:
∗
STB?
Response format:
<NR1>
Response example:
2