![NF EC1000SA Instruction Manual Download Page 206](http://html1.mh-extra.com/html/nf/ec1000sa/ec1000sa_instruction-manual_1684880206.webp)
6.3
Command Descriptions
EC1000SA
6-28
[SOURce:]VOLTage:LIMit:LOW
This sets the lower limit of the output voltage that can be set. However, a value that is equal to or
higher than the currently set cannot be specified. ([5, “Out of Limiter”] error occurs.)
For the 100 V range, the limit value setting ranges from
−
220.0 to
−
0.1 V, and for the 200 V range,
it ranges from
−
440.0 to
−
0.1 V.
Syntax:
[SOURce:]VOLTage:LIMit:LOW
<NR2>
Parameters:
−
220.0 to
−
0.1 (
−
440.0 to
−
0.1)
Command example:
VOLT:LIM:LOW
−
200
Query example:
VOLT:LIM:LOW?
Response format:
<NR2>
Response example:
−
200.0
[SOURce:]VOLTage:OFFSet[:IMMediate]
This sets the DC voltage when in AC+DC-INT, AC+DC-ADD, or AC+DC-SYNC mode. For the
100 V range, the DC output voltage setting ranges from
−
220.0 V to 220.0 V, and for the 200 V
range, it range, from
−
440.0 V to 440.0 V.
The DC output voltage cannot be set unless the mode is AC+DC-INT, AC+DC-ADD, or
AC+DC-SYNC mode. ([3, “Invalid in this mode”] error occurs.)
Syntax:
[SOURce:]VOLTage:OFFSet[:IMMediate]
<NR2>
Parameters:
−
220.0 to 220.0 (
−
440.0 to 440.0)
Command example:
VOLT:OFFS 10
Query example:
VOLT:OFFS?
Response format:
<NR2>
Response example:
10.0
[SOURce:]VOLTage:ADJust:OFFSet:AC
This sets the DC offset voltage adjustment value in mV when in AC-INT, AC-EXT, AC-ADD,
AC-SYNC mode. It ranges from
−
50.0 mV to 50.0 mV. (The setting range is common to 100 V
range and 200 V range.)
The DC offset voltage adjustment value cannot be set unless the mode is AC-INT, AC-EXT,
AC-ADD, AC-SYNC mode. ([3, “Invalid in this mode”] error occurs.)
Syntax:
[SOURce:]VOLTage:OFFSet:AC
<NR2>
Parameters:
−
50.0 to 50.0
Command example:
VOLT:ADJ:OFFS:AC 10
Query example:
VOLT:ADJ:OFFS:OFFS:AC?
Response format:
<NR2>
Response example:
10.0
Summary of Contents for EC1000SA
Page 1: ...NF Corporation PROGRAMMABLE AC DC POWER SOURCE EC1000SA Instruction Manual nbn Austria GmbH...
Page 2: ...DA00036667 003 PROGRAMMABLE AC DC POWER SOURCE EC1000SA Instruction Manual...
Page 236: ...6 9 Notes on Remote Control EC1000SA 6 58...
Page 260: ...8 5 Backup Battery EC1000SA 8 8...
Page 280: ...9 17 External Dimensions and Weight EC1000SA 9 20...