6.3
Command Descriptions
EC1000SA
6-27
[SOURce:]SEQuence:STEP
Specifies the target step number for sequence editing. The step number specification ranges from 1
to 255.
Syntax:
[SOURce:]SEQuence:STEP
<NR1>
Parameters:
1 to 255
Command example:
SEQ:STEP 1
Query example:
SEQ:STEP?
Response format:
<NR1>
Response example:
1
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]
AC output voltage is set using the Vrms unit (when the waveform is SIN or SQU) or the Vp-p unit
(when the waveform is ARB). For the 100 V range, output voltage setting ranges from 0.0 Vrms to
155.0 Vrms or from 0.0 Vp-p to 440.0 Vp-p, and for the 200 V range, it ranges from 0.0 Vrms to
310.0 Vrms or from 0.0 Vp-p to 880.0 Vp-p.
When in AC-EXT or AC+DC-EXT mode, the output AC voltage cannot be set or queried. ([3,
“Invalid in this mode”] error occurs.)
Syntax:
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]
<NR2>
Parameters:
0.0 to 155.0/0.0 to 440.0 (0.0 to 310.0/0.0 to 880.0)
Command example:
VOLT 100
Query example:
VOLT?
Response format:
<NR2>
Response example:
100.0
[SOURce:]VOLTage:LIMit:HIGH
This sets the upper limit of the output voltage that can be set. However, a value that is equal to or
lower than the currently set cannot be specified. ([5, “Out of Limiter”] error occurs.)
For the 100 V range, the limit value setting ranges from 0.1 V to 220.0 V, and for the 200 V range,
it ranges from 0.1 V to 440.0 V.
Syntax:
[SOURce:]VOLTage:LIMit:HIGH
<NR2>
Parameters:
0.1 to 220.0 (0.1 to 440.0)
Command example:
VOLT:LIM:HIGH 200
Query example:
VOLT:LIM:HIGH?
Response format:
<NR2>
Response example:
200.0
Summary of Contents for EC1000SA
Page 1: ...NF Corporation PROGRAMMABLE AC DC POWER SOURCE EC1000SA Instruction Manual nbn Austria GmbH...
Page 2: ...DA00036667 003 PROGRAMMABLE AC DC POWER SOURCE EC1000SA Instruction Manual...
Page 236: ...6 9 Notes on Remote Control EC1000SA 6 58...
Page 260: ...8 5 Backup Battery EC1000SA 8 8...
Page 280: ...9 17 External Dimensions and Weight EC1000SA 9 20...