9.6
Sequence Function
EC1000SA
9-11
9.6 Sequence Function
Output parameters can be changed abruptly or swept in sequence.
Signal source is INT only.
Sequence function settings are stored in memory with battery backup.
a) Number of sequences
1 sequence per operation mode (AC or AC+DC mode) and
output range (100 V or 200 V)
b) Number of steps
1 to 255 (within 1 sequence)
c) Step time
0.1 ms to 999.9999 s (resolution 0.1 ms)
d) Operations within step
Constant, keep, or linear sweep
e) Parameters
DC voltage, AC voltage, frequency, waveform, and
synchronous step output (2 bits)
f) Jump times
1 to 999 or infinite
g) Sequence control
Start:
Starts a sequence.
Stop:
Stops a sequence.
Hold:
Maintains current step output. Restarted when a sequence is
started.
Branch:
Branches to the specified step.