![NF EC1000SA Instruction Manual Download Page 155](http://html1.mh-extra.com/html/nf/ec1000sa/ec1000sa_instruction-manual_1684880155.webp)
5.4
MEASURE Screen
EC1000SA
5-25
5.4.4 Load power factor measurement
Calculated by effective power/apparent power
In the case that it exceeds the measurement range, “
− − − − −
” is displayed.
•
Measurement range:
0.00 to 1.00
•
Display resolution:
0.01
5.4.5 Load crest factor measurement
Calculated by (maximum values of
maximum current
and
minimum current
)/RMS
current.
In the case that it exceeds the measurement range, “
− − − − −
” is displayed.
•
Measurement range: 1.00 to 50.00
•
Display resolution: 0.01