![NF EC1000SA Instruction Manual Download Page 200](http://html1.mh-extra.com/html/nf/ec1000sa/ec1000sa_instruction-manual_1684880200.webp)
6.3
Command Descriptions
EC1000SA
6-22
[SOURce:]PHASe:CLOCk
This sets the sync signal source in the external synchronization mode (AC-SYNC or
AC+DC-SYNC). The sync signal source setting is either LINE (line synchronization) or EXT
(external synchronization). However, when output is on, the sync signal source cannot be changed.
([1, “Invalid with output on”] error occurs.)
Unless the mode is AC-SYNC or AC+DC-SYNC mode, the sync signal source setting cannot be
set. ([3, “Invalid in this mode”] error occurs.)
Syntax:
[SOURce:]PHASe:CLOCk
<CRD>
Parameters:
LINE/EXT
Command example:
PHAS:CLOC EXT
Query example:
PHAS:CLOC?
Response format:
<CRD>
Response example:
EXT
[SOURce:]PHASe[:IMMediate]
This sets the phase at the start of output. The phase setting range is 0.0
°
to 359.9
°
. However, when
output is on, the phase setting cannot be changed. ([1, “Invalid with output on”] error occurs.)
When in AC-EXT or AC+DC-EXT mode, the phase setting cannot be set. ([3, “Invalid in this
mode”] error occurs.)
Syntax:
[SOURce:]PHASe[:IMMediate]
<NR2>
Parameters:
0.0 to 359.9
Command example:
PHAS 90
Query example:
PHAS?
Response format:
<NR2>
Response example:
90.0
[SOURce:]SEQuence:CONDition?
This returns the sequence status. The response message is 0 (Idle mode), 1 (Run mode), or 2 (Hold
mode).
Syntax:
[SOURce:]SEQuence:CONDition?
Parameters:
None
Query example:
SEQ:COND?
Response format:
<NR1>
Response example:
0
Summary of Contents for EC1000SA
Page 1: ...NF Corporation PROGRAMMABLE AC DC POWER SOURCE EC1000SA Instruction Manual nbn Austria GmbH...
Page 2: ...DA00036667 003 PROGRAMMABLE AC DC POWER SOURCE EC1000SA Instruction Manual...
Page 236: ...6 9 Notes on Remote Control EC1000SA 6 58...
Page 260: ...8 5 Backup Battery EC1000SA 8 8...
Page 280: ...9 17 External Dimensions and Weight EC1000SA 9 20...