6-20
6. [
Tslope
]:
Time of pressure rising
7. [
Δ
Papnea
]: Pressure of apnea ventilation
8. [
fapnea
]:
Frequency of apnea ventilation
9. [
ATRC
]: Automatic tube resistance compensation
In APRV under NIV, you need to set the following ventilation parameters:
1. [
O2%
]:
Oxygen concentration
2. [
Phigh
]:
High pressure
3. [
Thigh
]:
Time of high pressure
4. [
Plow
]: Low
pressure
5. [
Tlow
]:
Time of low pressure
6. [
Tslope
]:
Time of pressure rising
7. [
Δ
Papnea
]: Pressure of apnea ventilation
8. [
fapnea
]:
Frequency of apnea ventilation
6.8 Change Alarm Limits
You can change the alarm limits for Paw, MV, ftot and TVe by pushing the Alarm Setup key
and selecting [
Alarm Limits
]. You can change the alarm limits for EtCO2 if your ventilator
is configured with CO2 module. You can also change alarm volume and apnea time. For
details, refer to
9 Alarms
.
6.9 Ventilate the Patient
WARNING
Before use on the patient, check that the oxygen concentration in the delivered gas
is consistent with the set value.
Select [
Start Ventilation
] when in Standby and the system begins to ventilate the patient
according to your settings.
Summary of Contents for SynoVent E5
Page 1: ...SynoVent E5 Ventilator Operator s Manual...
Page 2: ......
Page 24: ...2 2 2 2 Equipment Appearance 2 2 1 Front View 4 1 2 3 5 6 7 8 9 10 11 12...
Page 46: ...3 18 FOR YOUR NOTES...
Page 60: ...4 14 FOR YOUR NOTES...
Page 68: ...5 8 FOR YOUR NOTES...
Page 92: ...6 24 FOR YOUR NOTES...
Page 118: ...9 12 FOR YOUR NOTES...
Page 154: ...A 8 FOR YOUR NOTES...
Page 169: ...B 15 Temperature C Life 1192 exp 2 0 0239 Temperature Life of 20 C Figure...
Page 170: ...B 16 FOR YOUR NOTES...
Page 184: ...D 8 FOR YOUR NOTES...
Page 195: ......
Page 196: ...P N 046 000991 00 10 0...