8-7
NOTE
The P-V tool function is disabled in the following cases
:
patient type of Ped,
CPAP/PSV, NIV or apnea vent mode, within one minute after nebulization or
suction, within one minute after last P-V loop test.
The P-V tool function is not recommended when there is great leakge or when the
patient has spontaneous breathing. The relevent characteristic points the P-V tool
function provides are only for your reference.
1. Push the Menu key. Select [
Tool
] and select [
Procedure
].
2. Select
[
P-V Tools
] to access the P-V tools screen.
3. Read P-V tool related notes on the screen.
4. Select
[
Procedure
] and set Pstart, Flow, Pmax and Vlimit parameters on the Procedure
screen. The system acquires Tmax parameter value based on the calculation formula and
displays it on the menu screen.
Flow: gas delivery and expiration flows of the static P-V loop.
Pstart: starting pressure of the static P-V loop.
Pmax: maximum pressure which the static P-V loop can reach.
Vlimit: maximum volume which the static P-V loop can reach.
Tmax: maxium measurement time required for completing static P-V loop
measurement.
5. Select
[
Start
] and the system starts P-V tool measurement. If you select [
Stop Insp
]
during measurement, the system stops inspiration measurement test immediately and
starts expiration measurement. If you select [
Abort
] during measurement, the system
aborts measurement immediately.
Summary of Contents for SynoVent E5
Page 1: ...SynoVent E5 Ventilator Operator s Manual...
Page 2: ......
Page 24: ...2 2 2 2 Equipment Appearance 2 2 1 Front View 4 1 2 3 5 6 7 8 9 10 11 12...
Page 46: ...3 18 FOR YOUR NOTES...
Page 60: ...4 14 FOR YOUR NOTES...
Page 68: ...5 8 FOR YOUR NOTES...
Page 92: ...6 24 FOR YOUR NOTES...
Page 118: ...9 12 FOR YOUR NOTES...
Page 154: ...A 8 FOR YOUR NOTES...
Page 169: ...B 15 Temperature C Life 1192 exp 2 0 0239 Temperature Life of 20 C Figure...
Page 170: ...B 16 FOR YOUR NOTES...
Page 184: ...D 8 FOR YOUR NOTES...
Page 195: ......
Page 196: ...P N 046 000991 00 10 0...