WARNING
Nebulization yields the best performance at flow of 6 L/min. Nebulizers with other
flows can create significant errors in tidal volume and oxygen mix.
3.9 Install the CO2 Module
A. Fastening screws for CO2 module mounting plate
B. CO 2 module
C. Fastening screws for CO2 module
D. W ater trap
E. Sam pling line
F. CO 2/calibration connection
G. CO2 module connection line
1. Place the CO2 module mounting plate onto the ventilator’s handle. Then tighten the
fastening screws.