D-6
Nebulizer Valve
Failure
M The nebulizer valve is faulty. Restart the ventilator. If the
problem persists, contact your service personnel.
3-way Valve
Failure
M The 3-way valve is faulty. Restart the ventilator. If the
problem persists, contact your service personnel.
Insp. Temperature
Sensor Failure
H
The inspiratory temperature sensor is faulty. Restart the
ventilator. If the problem persists, contact your service
personnel.
Volume Limited
L
In pressure mode, the inspired tidal volume exceeds the
high alarm limit. Expiration is switched to when the high
limit is reached.
Check for leakage or change parameter. Decrease pressure
setting or increase tidal volume alarm limit.
Ctrl Module Init
Error
H
The control module selftest is passed at power-on. But
monitoring module system software’s sending parameter
configuration to monitoring module is failed. Restart the
ventilator. If the problem persists, contact your service
personnel.
Protection Module
Init Error
H
The protection module selftest is passed at power-on. But
protection module system software’s sending parameter
configuration to protection module is failed. Restart the
ventilator. If the problem persists, contact your service
personnel.
CO2
module
CO2 Init Error
H
An error occurs during CO2 module initialization.
The CO2 module is not installed properly or is faulty
Contact your service personnel.
CO2 Selftest Error H
An error occurs during CO2 module selftest.
Restart the ventilator. If the problem persists, contact your
service personnel.
CO2 Comm Stop
L
CO2 module malfunction or communication failure
occurs.
Restart the ventilator. If the problem persists, contact your
service personnel.
CO2 Comm Error
H
CO2 module communication failure. Restart the ventilator.
If the problem persists, contact your service personnel.
CO2 Sensor High
Temp
L
The temperature of the sensor assembly is too high
(>63
℃
). Check, stop using or replace the sensor.
CO2 Sensor Low
Temp
L
The temperature of the sensor assembly is too low (<5
℃
).
Check, stop using or replace the sensor.
CO2 High Airway
Pressure
L
The airway pressure is too high (>790 mmHg).
Check the pneumatic connections. Make sure that the
ventilator application site meets the environmental
specifications. Check for special sources that affect the
ambient pressure. Attempt to restart the ventilator.
Summary of Contents for SynoVent E5
Page 1: ...SynoVent E5 Ventilator Operator s Manual...
Page 2: ......
Page 24: ...2 2 2 2 Equipment Appearance 2 2 1 Front View 4 1 2 3 5 6 7 8 9 10 11 12...
Page 46: ...3 18 FOR YOUR NOTES...
Page 60: ...4 14 FOR YOUR NOTES...
Page 68: ...5 8 FOR YOUR NOTES...
Page 92: ...6 24 FOR YOUR NOTES...
Page 118: ...9 12 FOR YOUR NOTES...
Page 154: ...A 8 FOR YOUR NOTES...
Page 169: ...B 15 Temperature C Life 1192 exp 2 0 0239 Temperature Life of 20 C Figure...
Page 170: ...B 16 FOR YOUR NOTES...
Page 184: ...D 8 FOR YOUR NOTES...
Page 195: ......
Page 196: ...P N 046 000991 00 10 0...