B-7
P0.1 -20
to
0
0.1
cmH
2
O
PEEPi
0 to 120
0.1
cmH
2
O
FiO
2
15 to 100
1
%
B.7 Ventilator Accuracy
Control accuracy
O
2
%
±3 vol.% or ±5% of setting, whichever is greater
TV
±10 mL or ±10% of setting, whichever is greater
f
1 bpm to 100 bpm: ±1 bpm
Other range: ±2% of setting
fSIMV ±1
bpm
Tinsp
±0.1s or ±10% of setting, whichever is greater
I: E
2: 1 to 1: 4: ±10% of setting
Other range: not defined.
Flow
±1 L/min or ±20% of setting, whichever is greater
Tslope ±0.2s
Plimit ±2.0
cmH
2
O or ±10% of setting, whichever is greater
PEEP ±2.0
cmH
2
O or ±10% of setting, whichever is greater
Pinsp ±2.0
cmH
2
O or ±10% of setting, whichever is greater
△
Psupp ±2.0
cmH
2
O or ±10% of setting, whichever is greater
Phigh ±2.0
cmH
2
O or ±10% of setting, whichever is greater
Plow ±2.0
cmH
2
O or ±10% of setting, whichever is greater
Thigh
±0.2s or ±10% of setting, whichever is greater
Tlow
±0.2s or ±10% of setting, whichever is greater
Trigger ±1.0
cmH
2
O or ±10% of setting, whichever is greater or,
±1.0 L/min or ±20% of setting, whichever is greater
△
int.PEEP ±2.0
cmH
2
O or ±20% of reading whichever is greater
Exp% ±10%
fapnea ±1
bpm
△
Papnea ±2.0
cmH
2
O or ±10% of setting, whichever is greater
Monitoring accuracy
Ppeak
±(2 cmH
2
O + 4 % of the actual reading)
Pplat
Pmean
PEEP
Summary of Contents for SynoVent E5
Page 1: ...SynoVent E5 Ventilator Operator s Manual...
Page 2: ......
Page 24: ...2 2 2 2 Equipment Appearance 2 2 1 Front View 4 1 2 3 5 6 7 8 9 10 11 12...
Page 46: ...3 18 FOR YOUR NOTES...
Page 60: ...4 14 FOR YOUR NOTES...
Page 68: ...5 8 FOR YOUR NOTES...
Page 92: ...6 24 FOR YOUR NOTES...
Page 118: ...9 12 FOR YOUR NOTES...
Page 154: ...A 8 FOR YOUR NOTES...
Page 169: ...B 15 Temperature C Life 1192 exp 2 0 0239 Temperature Life of 20 C Figure...
Page 170: ...B 16 FOR YOUR NOTES...
Page 184: ...D 8 FOR YOUR NOTES...
Page 195: ......
Page 196: ...P N 046 000991 00 10 0...