6-1
6
Start Ventilation
6.1 Turn on the System
1. Plug the power cord into the AC power outlet. Make sure that the AC power LED is lit.
2. Push and hold the
key.
3. The alarm LED flashes yellow and red once in turn and then the speaker and buzzer give
a self-test sound.
4. A start-up screen and self-test progress bar appear. Then the standby screen is displayed.
6.2 Preoperative Test
6.2.1 AC Power and Battery Power Source Switch Test
1. Push
the
key to start the ventilator.
2. Remove AC power. Make sure that the AC power LED is extinguished, the battery
indicator flashes, and the low level alarm [
Battery in Use
] is triggered.
3. Reconnect AC power and operate the ventilator on AC mains. Make sure that the AC
power LED is lit, the battery indicator stops flashing and remains illuminated, and the
alarm [
Battery in Use
] disappears.
4. Push
the
key again to turn off the ventilator.
6.2.2 Pipeline Tests
6.2.2.1 O2 Pipeline Test
1. Connect an O2 pipeline supply.
2. Connect the test splint lung.
3. Push
the
key to start the ventilator.
4. Select adult ventilation mode and make the ventilator start ventilation. Make sure that
the ventilator ventilates properly.
5. Disconnect the O2 pipeline supply.
6. The high level alarm [
O2 Supply Pressure Low
] is triggered with the decrease in O2
pressure.
Summary of Contents for SynoVent E5
Page 1: ...SynoVent E5 Ventilator Operator s Manual...
Page 2: ......
Page 24: ...2 2 2 2 Equipment Appearance 2 2 1 Front View 4 1 2 3 5 6 7 8 9 10 11 12...
Page 46: ...3 18 FOR YOUR NOTES...
Page 60: ...4 14 FOR YOUR NOTES...
Page 68: ...5 8 FOR YOUR NOTES...
Page 92: ...6 24 FOR YOUR NOTES...
Page 118: ...9 12 FOR YOUR NOTES...
Page 154: ...A 8 FOR YOUR NOTES...
Page 169: ...B 15 Temperature C Life 1192 exp 2 0 0239 Temperature Life of 20 C Figure...
Page 170: ...B 16 FOR YOUR NOTES...
Page 184: ...D 8 FOR YOUR NOTES...
Page 195: ......
Page 196: ...P N 046 000991 00 10 0...