5-4
5.6.2 Set Paw Unit
1. Push the Menu key. Select [
System
] and then [
Unit
].
2. Select
[
Paw Unit
] and toggle between [
cmH2O
], [
hPa
], and [
mbar
].
5.6.3 Set CO2 Unit
1. Push the Menu key. Select [
System
] and then [
Unit
].
2. Select
[
CO2 Unit
] and toggle between [
mmHg
], [
kPa
], and [
%
].
5.7 Turning on/off O2% Monitoring
1. Push the Menu key. Select [
System
] and then [
Ventilator
].
2. Select
[
O2% Monitoring
] and toggle between [
ON
] and [
OFF
]. When [
ON
] is selected,
oxygen concentration of patient’s inspired gas can be monitored. It needs about 30s after
powering on the ventilator till the point that the oxygen concentration monitoring
function enters full accuracy mode. You can set [
O2% Monitoring
] to [
OFF
] if oxygen
concentration monitoring function accompanying the ventilator is not needed. In this
case, [
O2 Monitoring Off
] is prompted on the screen.
5.8 Select Tinsp/I:E
1. Push the Menu key. Select [
System
] and then [
Ventilator
].
2. Select
[
Tinsp/I:E
] and toggle between [
Tinsp
] and [
I:E
]. Based on your Tinsp/I:E
selection, corresponding Tinsp or I/E ventilation setting parameters are adopted for
V-A/C, P-A/C and PRVC modes.
5.9 Set TV/f Source
1. Push the Menu key. Select [
System
] and then [
Ventilator
].
2. Select
[
TV/f Source
] and toggle among [
Height&Gender
], [
IBW
] and [
Patient Type
].
When the ventilator is used on a new patient, the system sets the default TV, f or fapnea
values based on the TV/f source setting.
5.10 Set Sigh
1. Push the Menu key. Select [
System
] and then [
Ventilator
].
2. Select
[
Sigh
] and toggle between [
TV
] and [
Δ
int.PEEP
].
Summary of Contents for SynoVent E5
Page 1: ...SynoVent E5 Ventilator Operator s Manual...
Page 2: ......
Page 24: ...2 2 2 2 Equipment Appearance 2 2 1 Front View 4 1 2 3 5 6 7 8 9 10 11 12...
Page 46: ...3 18 FOR YOUR NOTES...
Page 60: ...4 14 FOR YOUR NOTES...
Page 68: ...5 8 FOR YOUR NOTES...
Page 92: ...6 24 FOR YOUR NOTES...
Page 118: ...9 12 FOR YOUR NOTES...
Page 154: ...A 8 FOR YOUR NOTES...
Page 169: ...B 15 Temperature C Life 1192 exp 2 0 0239 Temperature Life of 20 C Figure...
Page 170: ...B 16 FOR YOUR NOTES...
Page 184: ...D 8 FOR YOUR NOTES...
Page 195: ......
Page 196: ...P N 046 000991 00 10 0...