9-10
3. After the CO2 test module finishes the preheating and enters in the working status,
connect the CO2 of 3% to 7% to the simple port of the sidestream CO2 module or the
airway adapter of the mainstream CO2 module. Set the EtCO2 high alarm limit to be
lower than the concentration of the standard gas.
4. Check
if
[
EtCO2 Too High
] alarm is triggered.
9.11.12 EtCO2 Too Low Alarm
1. Connect the test lung to the ventilator, and start ventilation.
2. Connect the CO2 test module to the ventilator, and set the CO2 test module to working
condition.
3. After the CO2 test module finishes the preheating and enters in the working status,
connect the standard CO2 of 3% to 7% to the simple port of the sidestream CO2 module
or the airway adapter of the mainstream CO2 module. Set the EtCO2 high alarm limit to
be higher than the concentration of the standard gas.
4. Check
if
[
EtCO2 Too Low
] alarm is triggered.
9.12 Alarm Logbook
For alarm logbook, the system provides up to 500 events which are stored in chronological
order. When a new event occurs after 500 events are already stored, the earliest event will be
overwritten by the new one. Each alarm log includes the ventilation mode, ventilation type
(NIV or non-NIV), patient type (adult or pediatric), operating mode (standby or non-standby
mdoe), and monitored parameters.
To access the Alarm Logbook window, push the Alarm Setup key and select [
Alarm
Logbook
]. In the Alarm Logbook window, you can:
1. Select
[
Scroll
] to view the alarm events item by item.
2. Select
[
Alarm Level
] and select the desired level of alarms to be displayed. The options
include all level, high level, medium level, and low level.
NOTE
After the ventilator is started up, if [Same Patient] is applied, the system will store
the previous alarm log and continue recording. If [New Patient] is applied, the
system will clear the previous alarm log and record anew.
Summary of Contents for SynoVent E5
Page 1: ...SynoVent E5 Ventilator Operator s Manual...
Page 2: ......
Page 24: ...2 2 2 2 Equipment Appearance 2 2 1 Front View 4 1 2 3 5 6 7 8 9 10 11 12...
Page 46: ...3 18 FOR YOUR NOTES...
Page 60: ...4 14 FOR YOUR NOTES...
Page 68: ...5 8 FOR YOUR NOTES...
Page 92: ...6 24 FOR YOUR NOTES...
Page 118: ...9 12 FOR YOUR NOTES...
Page 154: ...A 8 FOR YOUR NOTES...
Page 169: ...B 15 Temperature C Life 1192 exp 2 0 0239 Temperature Life of 20 C Figure...
Page 170: ...B 16 FOR YOUR NOTES...
Page 184: ...D 8 FOR YOUR NOTES...
Page 195: ......
Page 196: ...P N 046 000991 00 10 0...